Other
  • 13 Jul 2020
  • 1 Minute to read
  • Contributors
  • Dark
    Light
  • PDF

Other

  • Dark
    Light
  • PDF

Article Summary

Is there default values expected to be seen if a sensor or sensor circuit is shorted or open?

The device will display "-----" if the sensor is not responding correctly.

What liquid is used in the probe bottles?

Propylene Glycol.

Are there installation instructions for sensor placement?

It is very dependent on equipment and application. In general, for best uniformity of refrigeration, it is recommended to use two probes per refrigerator, one in the back nearest to the compressor/coil locations, and the other at the furthest point from that, nearest to the door. Some VFC guidance directs that a single probe be placed in the center of a refrigerator.

What is the total number of measurement points stored in each device?

672 readings per channel (7 days@ 15 min interval) and 100 stored alarm events.

How does TraceableLIVE® comply to the 21 CFR Part 11 Standard?

Plese see our 21 CFR Part 11 documentation for details.


Was this article helpful?